| System Requirements | ||||
| SYSTEM2/10EH | CELLBAG2L and 10L | Integral temperature control.
No incubator needed.
Integral airpump with mass flow meter. Space needed: 19" wide x 13" deep x 8" high 110/220 VAC. |
||
| SYSTEM20/50P
|
CELLBAG2L, 10L, 20L, 22L and 50L | Pneumatic operation. Incubator
or hot room required. Optional temperature control (HEATER20 or 50) for benchtop operation. Space needed: 31" wide x 28" deep x 13" high. Needs 2 to 6 bar (30 to 90 psig) compressed air (@ 0.2 scfm). No electrical power required. Built-in pressure regulator." 24 volts DC 100mA for aeration pump (AIRPUMP20). 115 VAC power adapter provided. |
||
| SYSTEM20/50EH
|
CELLBAG2L, 10L, 20L, 22L and 50L | Electric operation with integral
temperature control. No incubator required. Space needed: 31" wide x 28" deep x 13" high 110/220 50/60 Hz autoswitching for rocker and heater. |
||
| SYSTEM200EH | CELLBAG100L and 200L | Floor standing unit mounted
on casters. Integral temperature control. No incubator needed. Space needed: 73" wide x 43" width x 44" height. 200-240 VAC (15 amp) 3 phase. NEMA L2130 Plug. |
||
| SYSTEM500/1000EH | CELLBAG500L and 1000L |
Floor standing unit mounted on casters. Integral temperature control. No incubator needed. Space needed 89" wide x 91" length x 63" height. 208/240 VAC (30 amp) 3 phase. | ||