Question

Solution

What is range of culture volume possible in the SYSTEM500/1000?

SYSTEM500/1000 can accommodate a CELLBAG1000L which has a maximum liquid volume of 500 liters. Alternatively, you can put two CELLBAG500L side-by-side giving you two cultures of up to 250 liters each.

What utilities do I need for the SYSTEM500/1000?

You will need 208/240 VAC (3 phase). Also CO2 and O2 depending on your applications. Gas consumption rate is very low, and a standard cylinder will last over a month, alternatively, you may use our O2GEN.

I am not sure that the SYSTEM500/1000 will work for my application?

We have several of SYSTEM500/1000s in place. The first units were delivered in late 2002. The technology is well-proven, however we recommend you gain some experience with a smaller SYSTEM20/50 or SYSTEM200 first, then scale-up is usually trivial.

How much space does it take?

The footprint is about 7.5 x 8 feet (2.3 x 2.4 meters).